Gw-instek PSM-Series Programming Manual Manuale Utente Pagina 17

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 36
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 16
PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
31
Examples:
:CURRent? returns +1.20000000E-02 if the current limit setting is 0.0120
amps(PSM-2010).
:CURRent? MIN returns +0.00000000E+00(PSM-2010)..
:CURRent? MAX returns +2.06000000E+01 (PSM-2010).
:VOLTage
Function:
Set the output voltage value.
Syntax:
:VOLTage {<voltage>|MIN|MAX|UP|DOWN}
1) <voltage> the range is from 0.0000 through 20.6000 (PSM-2010).
Please refer to Table1-1, Table1-2 and Table1-3.
2) MIN is for 0.0000 while MAX for 20.6000(PSM-2010).
Please refer to
Table1-1, Table1-2 and Table1-3.
3) UP is for the voltage with one value increasing while down for the
voltage with one value decreasing. The value is set by the command of
VOLTage:STEP.
Examples:
:VOLTage 2.0 sets the voltage limit to 2.0000 volts (PSM-2010).
:VOLTage MAX sets the voltage limit to 20.6000 volts (PSM-2010).
:VOLTage:STEP 0.1;VOLTage UP is for Voltage to increase 0.1 volts.
:VOLTage?
Function:
Query the output voltage value.
PSM SERIES PROGRAMMABLE POWER SUPPLY
PROGRAMMER MANUAL
32
Syntax:
:VOLTage? [MIN|MAX]
Returns:
Scientific mark
Examples:
:VOLTage? returns +1.20000000E-02 if the voltage limit setting is 0.0120
volts(PSM-2010).
:VOLTage? MIN returns +0.00000000E+00(PSM-2010).
:VOLTage? MAX returns +2.06000000+01(PSM-2010).
:CURRent:STEP
Function:
Set or query the current step size.
Syntax:
:CURRent:STEP {<Decimal number>|DEFault}
:CURRent:STEP? [DEFault]
Examples:
1) :CURRent:STEP 1.0 sets the step size to 1.0A for current
programming with the CURRent:UP and CURRent:DOWN
command .The immediate current level increases or decreases by the
value of the step size.
2) :CURRent:STEP DEF sets the step size to minimum resolution
0.5mA(PSM-2010).
3) :CURRent:STEP? returns the present current step size.
4) :CURRent:STEP? DEF returns the minimum resolution step size
0.5mA(PSM-2010).
Vedere la pagina 16
1 2 ... 12 13 14 15 16 17 18 19 20 21 22 ... 35 36

Commenti su questo manuale