REMOTE CONTROL
103
Parameter
<ampl>
<NRf> Power or voltage in the current Y-
axis unit.
Return parameter
<NR3>
Example
:DISP:TRAC:Y:DLIN -5.0e+1
:DISPlay[:WINDow]:TRACe:Y:DLINe:STATe
Set
Query
Description
Turns the display line on/off or queries its state.
Syntax
:DISPlay[:WINDow]:TRACe:Y:DLINe:STATe
{OFF|ON|0|1}
Query Syntax
:DISPlay[:WINDow]:TRACe:Y:DLINe:STATe?
Parameter
0
Turn display line off.
1
Turn display line on.
OFF
Turn display line off.
ON
Turn display line on.
Return parameter
0
The display line is off.
1
The display line is on.
Example
:DISP:TRAC:Y:DLIN:STAT ON
:DISPlay[:WINDow]:TRACe:Y[:SCALe]:
NRLevel
Set
Query
Description
Sets or queries the normalized reference level for
the TG option.
Syntax
:DISPlay[:WINDow]:TRACe:Y[:SCALe]:NRLevel <ampl>
Query Syntax
:DISPlay[:WINDow]:TRACe:Y[:SCALe]:NRLevel?
Parameter
<ampl>
<NRf> Power or voltage in the current Y-
axis unit.
Return parameter
<NR3>
Example
:DISP:TRAC:Y:NRLevel 5 dBm