Programmable DC Electronic Load PEL-2000 Series PROGRAMMING MANUAL GW INSTEK PART NO. 82EL-20040IA1 ISO-9001 CERTIFIED MANUFACTURER

Pagina 2

PEL-2000 Programming Manual 10 Functionality check Run this query command via the terminal. *idn? This should return the Manufacturer, Model numbe

Pagina 3 - Table of Contents

PEL-2000 Programming Manual 100 Example :PROG:ONT 10S Set the on-time for the current program number to 10 seconds. Query Syntax :PROGram:ONTime?

Pagina 4

COMMAND DETAILS 101 :PROGram:RUN All Channel Command Description Runs the current program number when set to on, and when set to off will allow a

Pagina 5 - INTERFACE OVERVIEW

PEL-2000 Programming Manual 102 0.0~119.9 Returns the P/F-Time for the program sequence. Query Example :PROG:PFTime? 5 The P/F-Tme is 5 seconds.

Pagina 6 - Interface

COMMAND DETAILS 103 Resistance Subsystem :RESistance[:STATic]:L1/L2 ... 103 :RESistance[:STATic]:RISE/FALL

Pagina 7

PEL-2000 Programming Manual 104 Example :RES:L1 10 Sets CR static mode A Value to 10 ohms, depending on the specific range Query Syntax :RESistanc

Pagina 8

COMMAND DETAILS 105 Return Parameter <NR2> [MAX|MIN] Slew rate 1 Unit =1 amp/uS Returns the slew rate. MAX/MIN Returns the allowable maxim

Pagina 9

PEL-2000 Programming Manual 106 BVALue 20 Sets B Value to 20 ohms. (Low Range only) AVALue MIN Sets A Value to the minimum level for the specifi

Pagina 10

COMMAND DETAILS 107 Example :RES:STAT:LOW:RISE 0.1 Sets the rising slew rate to 0.1A/uS. Query Syntax :RESistance:STATic:LOW:RISE/FALL? [MIN|MAX]

Pagina 11 - GPIB Interface Configuration

PEL-2000 Programming Manual 108 Query Syntax :RESistance:STATic:HIGH:AVALue/BVALue? [MAX|MIN] Return Parameter <NR2> [MAX|MIN] Resistance

Pagina 12

COMMAND DETAILS 109 Query Example :RES:STAT:HIGH:RISE? MIN 0.8000 For high range CR mode, the minimum value for the rising slew rate is 0.8 A/uS f

Pagina 13 - COMMAND OVERVIEW

INTERFACE OVERVIEW 11 GPIB Interface Configuration Panel operation 1. Press the Shift Key then the Help key to access the Utility menu. UTILITYH

Pagina 14 - FETCh:VOLTage?

PEL-2000 Programming Manual 110 :RESistance:DYNamic:LOW:RISE/FALL Channel Specific Command Description Sets the low range rising/falling slew rat

Pagina 15

COMMAND DETAILS 111 Parameter <NRf+>[S] Time T1/T2 T1/T2 0.1S Sets the T1/T2 time to 0.1 seconds. T1/T2 1 Sets T1/T2 to 1 second. T1/T2 M

Pagina 16

PEL-2000 Programming Manual 112 L1/L2 MIN Sets L1 or L2 to the minimum level for the specific channel. L1/L2 MAX Sets L1 or L2 to the maximum le

Pagina 17

COMMAND DETAILS 113 Query Syntax :RESistance:DYNamic:HIGH:FALL? [MIN|MAX] Return Parameter <NR2> [MAX|MIN] Slew rate 1 Unit=1 amp/uS Return

Pagina 18

PEL-2000 Programming Manual 114 Query Example :RES:DYN:HIGH:T1? 2.5 :RES:DYN:LOW:T1? MIN 0.000025 Returns the T1 time of 2.5 seconds. Returns the

Pagina 19

COMMAND DETAILS 115 RUN Subsystem :RUN ... 115 :RUN All Channel Comm

Pagina 20

PEL-2000 Programming Manual 116 SHOW Subsystem :SHOW[:DISPlay] dual channel ... 116 :SHOW[:DISPlay] single channe

Pagina 21

COMMAND DETAILS 117 :SHOW[:DISPlay] single channel Channel Specific Command (Single channel module) Description Sets the display mode of the loa

Pagina 22

PEL-2000 Programming Manual 118 SPECIFICATION Subsystem :SPECification:UNIT ... 118 :SPECificat

Pagina 23 - Global

COMMAND DETAILS 119 :SPECification[:PASS]? Channel Specific Command Description Displays if the Go/NoGo (specification) limit has passed/failed f

Pagina 24 - COMMAND DETAILS

PEL-2000 Programming Manual 12 GPIB constraints  Maximum 15 devices altogether, 20m cable length, 2m between each device  Unique address assig

Pagina 25 - COMMAND DETAILS

PEL-2000 Programming Manual 120 :SPECification:VOLTage:H/L/C Channel Specific Command Description Sets the high(H), low(L) and center(C) Go/NoGo

Pagina 26 - Common commands

COMMAND DETAILS 121 Return Parameter <NR2> Go/NoGo voltage 1 unit = 1 amp Returns the limit current Query Example :SPEC:CURR:H? 5.120 The

Pagina 27

PEL-2000 Programming Manual 122 STATUS Subsystem :STATus:CHANnel:CONDition? ... 122 :STATus:CHANnel:ENABle ...

Pagina 28 - Manufacturer

COMMAND DETAILS 123 :STATus:CHANnel:ENABle Channel Specific Command Description Sets which events are enabled in the Channel Status Enable regis

Pagina 29

PEL-2000 Programming Manual 124 Return Parameter <NR1> Event <NR1> Event 1 OC 16 OT 2 OV 32 G/N 4 OP 64 UVP 8 RV 128~65535 Not Use

Pagina 30

COMMAND DETAILS 125 Query Syntax :STATus:CHANnel:NTRansition/PTRansition? <NR1> Return Parameter <NR1> Event <NR1> Event 1 OC 1

Pagina 31

PEL-2000 Programming Manual 126 Query Example :STAT:CSUM:ENAB? 4 Only the events from channel 3 can set the CSUM bit in the Status Byte Register.

Pagina 32 - Abort Subsystem

COMMAND DETAILS 127 Query Example :STAT:QUES:COND? 2 OV (overvoltage) error. :STATus:QUEStionable:ENABle Channel Specific Command Description Se

Pagina 33 - Channel Subsystem

PEL-2000 Programming Manual 128 Query Syntax :STATus:QUEStionable[:EVENt]? <NR1> Return Parameter <NR1> Event <NR1> Event 1 OC

Pagina 34 - :CHANnel:SYNCon

COMMAND DETAILS 129 Example :STAT:QUES:NTR 5 OC (Bit 1) and OP (Bit 3) are set as negative transitions. :STAT:CHAN:PTR 2 OV (Bit 2) is s

Pagina 35 - :CHANnel:DISPlay

COMMAND OVERVIEW 13 COMMAND OVERVIEW The Command overview chapter lists all the PEL-2000 commands and command queries .The command syntax section

Pagina 36 - :CHANnel:MEMo

PEL-2000 Programming Manual 130 Voltage Subsystem :VOLTage:L1/L2 ... 130 :VOLTage:RECall

Pagina 37

COMMAND DETAILS 131 MAX/MIN Returns the allowable maximum and minimum. Query Example :VOLT:L1? 5 :VOLT:L1? MAX 81.6000 A Value is set to 5 volts.

Pagina 38 - CONFIGURE Subsystem

PEL-2000 Programming Manual 132 MAX Sets the voltage to the maximum value for the channel Example :VOLT:AVAL 10V Sets A Value to 10 volts for the

Pagina 39 - :CONFigure:VOLTage:RANGe

COMMAND DETAILS 133 MAX Sets the current limit to the maximum value for the channel Example :VOLT:CURR 1A Sets the current limit to 1 Amp for the

Pagina 40 - :CONFigure:AUTO:LOAD

PEL-2000 Programming Manual 134 Return Parameter <NR1> Response Time 0 Slow 1 Fast Query Example :VOLT:MODE? 0 The specific channel is se

Pagina 41 - :CONFigure:SOUND

COMMAND DETAILS 135 Power Subsystem :POWer:L1/L2 ... 135 :POWer:CURRent ...

Pagina 42 - :CONFigure:REMote

PEL-2000 Programming Manual 136 1 unit = 1 watt Returns the power of the A Value (L1) or B Value (L2). MAX/MIN Returns the allowable maximum and

Pagina 43 - :CONF:PROT:CURR:STAT?

COMMAND DETAILS 137 Query Example :POW:CURR? 7.0 Returns the current limit for the specific channel. :POWer:RECall Channel Specific Command Desc

Pagina 44

PEL-2000 Programming Manual 138 AVALue MAX Sets A Value to the maximum Level for the specific channel. Example :POWer:LOW:AVAL 1 Sets A Value to

Pagina 45

COMMAND DETAILS 139 Query Syntax :POW:LOW:CURRent? [MAX|MIN] Return Parameter <NR2> [MAX|MIN] Current 1 unit = 1 amp Returns the current

Pagina 46

PEL-2000 Programming Manual 14 Command forms Commands and queries have two different forms, long and short. The command syntax is written with the

Pagina 47

PEL-2000 Programming Manual 140 Query Example :POWer:HIGH:BVAL? AX 2 Returns the maximum power allowed for the channel/range. :POWer:HIGH:CURRent

Pagina 48 - :CONFigure:RESPonse

COMMAND DETAILS 141 SYSTEM Subsystem :SYSTem:ERRor? ... 141 :SYSTem:VERSion?...

Pagina 49 - :CONFigure:GROup:UNITs

PEL-2000 Programming Manual 142 :SYSTem:SETup System Command Description Sets or returns the system setup for the current settings using block da

Pagina 50 - :CONFigure:GROup:MODE

COMMAND DETAILS 143 Memory Subsystem :MEMory:SAVE:PREset ... 143 :MEMory:SAVE:PROGram ...

Pagina 51 - Utility Subsystem

PEL-2000 Programming Manual 144 Example :MEM:SAVE:PROG 100 Saves the channel to Memory M100. :MEMory:SAVE:ALLPreset All Channels Description S

Pagina 52 - :UTILity:SOUNd

COMMAND DETAILS 145 Example :MEM:REC:PRE 0 Recalls the preset settings from (P0). :MEMory:RECall:PROGram Channel Specific Command Descrip

Pagina 53 - :UTILity:REMote:MODE

PEL-2000 Programming Manual 146 Example :MEM:REC:SET 1 Recalls the setup data from S1 (applicable to all channels). Example :MEM:REC:100

Pagina 54

COMMAND DETAILS 147 Return Parameter <block data> Returns the program data as block data. Query Example :MEM:FILE:PROG? #<digits><

Pagina 55 - :UTILity:FRAMe

PEL-2000 Programming Manual 148 Return Parameter <block data> Returns the sequence data as block data. Query Example :MEM:FILE:SEQuence? #&

Pagina 56 - Current Subsystem

COMMAND DETAILS 149 SEQuence Subsystem :SEQuence:EDIT:POINt ... 149 :SEQuence:END ...

Pagina 57 - :CURRent:STATic:L1/L2

COMMAND OVERVIEW 15 Parameter Type Description Example <Boolean> Boolean logic 0, 1 <NR1> integers 0, 1, 2, 3 <NR2> decimal

Pagina 58 - :CURRent:STATic:RISE/FALL

PEL-2000 Programming Manual 150 Query Example :SEQ:EDIT:POIN? 3 The current point is 3. :SEQuence:END Channel Specific Command Description Sets

Pagina 59

COMMAND DETAILS 151 1 unit = 1 ohm Returns the resistance value. MAX/MIN Returns the allowable maximum and minimum. Query Example :SEQ:POIN:RES?

Pagina 60 - :CURR:STAT:LOW:RISE .001

PEL-2000 Programming Manual 152 1.2 1.2A/us 1.2 A/us 1.2A/us MAX/MIN Maximum or minimum value. Example :SEQ:POIN:RISE .3 A Sets the rising sle

Pagina 61

COMMAND DETAILS 153 Return Parameter <NR2> Point 0.0001~60000 Returns the point duration time. MAX/MIN Returns the allowable maximum and

Pagina 62 - :CURRent:DYNamic:L1/L2

PEL-2000 Programming Manual 154 Parameter OFF/0 Off ON/1 On Example :SEQ:END:LOAD 0 On End Load = Off. Query Syntax :SEQuence:END:LOAD? Return Pa

Pagina 63 - :CURRent:DYNamic:RISE/FALL

COMMAND DETAILS 155 Query Example :SEQ:VOLT:RANG? 500 Returns the voltage range. In this case high for the PEL-2041. :SEQuence:LOOP:STARt Channel

Pagina 64 - :CURRent:DYNamic:T1/T2

PEL-2000 Programming Manual 156 Query Syntax :SEQuence:CHANnel:TIME? Return Parameter <NR1> Point 1~max channels Returns the channel durat

Pagina 65 - :CURRent:DYNamic:LOW:L1/L2

COMMAND DETAILS 157 Example :SEQ:TRIG:IN 0 Trigger IN is on. Query Syntax :SEQuence:TRIGger:IN? Return Parameter <NR1> Trigger IN 0 Off 1

Pagina 66

PEL-2000 Programming Manual 158 Syntax :SEQuence:TRIGger:IN:CHANnel <NR1> Parameter <NR1> Channel <NR1> Channel 1 CH1 32 CH 6

Pagina 67 - :CURRent:DYNamic:HIGH:L1/L2

COMMAND DETAILS 159 GLOBal Subsystem :GLOBal:CONFigure:VOLTage:RANGe ... 159 :GLOBal:LOAD:SHORt ...

Pagina 68

PEL-2000 Programming Manual 16 List of Commands in Functional Order Common Commands *CLS ...

Pagina 69 - :CURRent:DYNamic:HIGH:T1/T2

PEL-2000 Programming Manual 160 Syntax :GLOBal:MODE {CCL|CCH|CCDL|CCDH|CRL|CRH|CRDL|CRDH|CV|CPL|CPH} Parameter CCL CC static mode, low range CCH

Pagina 70

COMMAND DETAILS 161 Command Error Codes Description The PEL has a number of specific error codes. Use the SYSTem:ERRor command to recall the erro

Pagina 71 - FETCH Subsystem

PEL-2000 Programming Manual 162 STATUS REGISTERS To program the PEL-200 Series effectively, the Status Register structure needs to be understood.

Pagina 72 - :FETCh:ALLVoltage?

STATUS REGISTERS 163 BIT WEIGHT1 2 4 8OR16CONDITION1 2 4 8PTR/NTR1 2 4 8EVENT1 2 4 8ENABLEOC OV OP RV OTProtection0 1 2 3 4Bit positionBIT WEIGHT1

Pagina 73 - :FETCh:ALLPower?

PEL-2000 Programming Manual 164 Channel Status Description Each channel has a dedicated Channel Status Register group. These registers show if an

Pagina 74

STATUS REGISTERS 165 G/N The Go/NoGo bit is set when Go/NoGo limits have been exceeded, when Go/NoGo SPEC has been enabled. UVP If the under vol

Pagina 75 - LOAD Subsystem

PEL-2000 Programming Manual 166 Channel Summary Registers834221101287646325164Bit weightBit PositionEVENTENABLECH4 CH3 CH2 CH1CH8 CH7 CH6 CH5CH4

Pagina 76 - :LOAD:PROTection?

STATUS REGISTERS 167 UVP Under Voltage Protection Condition Register The Questionable Status Condition Register indicates the status of the electr

Pagina 77 - :LOAD:TIME?

PEL-2000 Programming Manual 168 Standard Event Status Description The Standard Event Status Registers indicate any programming errors that occur.

Pagina 78 - :LOAD:TYPE

STATUS REGISTERS 169 CME The Command Error bit is set when a syntax error has occurred. The CME bit can also be set when a <GET> command is

Pagina 79

COMMAND OVERVIEW 17 Configure :CONFigure:VOLTage:ON ... 38 :CONFigure:VOLTage:RANGe ...

Pagina 80 - Measure Subsystem

PEL-2000 Programming Manual 170 MSS & RQS The Master Summary Status is used with the *STB? query. When the *STB? query is read the MSS bit is

Pagina 81 - :MEASure:INPut

PEL-2000 Programming Manual 18 Current :CURRent:STATic:RECall ... 56 :CURRent:STATic:L1/L2 ...

Pagina 82 - :MEASure:ALLVoltage?

COMMAND OVERVIEW 19 Measure :MEASure:VOLTage? ... 80 :MEASure:CURRent? ...

Pagina 83 - :MEASure:ALLPower?

September 2012 edition This manual contains proprietary information which is protected by copyright. All rights are reserved. No part of this manual

Pagina 84

PEL-2000 Programming Manual 20 Program :PROGram:FILE ... 94 :PROGram:SEQuence ...

Pagina 85 - MODE Subsystem

COMMAND OVERVIEW 21 Specification :SPECification:UNIT ... 118 :SPECification[:PASS]? ...

Pagina 86

PEL-2000 Programming Manual 22 System :SYSTem:ERRor? ... 141 :SYSTem:VERSion? ...

Pagina 87 - OCP Test Automation Commands

COMMAND OVERVIEW 23 Global :GLOBal:CONFigure:VOLTage:RANGe ... 159 :GLOBal:LOAD:SHORt ...

Pagina 88 - :OCP:CHANnel:STARt

PEL-2000 Programming Manual 24 COMMAND DETAILS The Command details chapter shows the detailed syntax, equivalent panel operation, and example for

Pagina 90 - :OCP:CHANnel:STEP:TIME

PEL-2000 Programming Manual 26 Common commands *CLS ... 26 *ESE ...

Pagina 91 - :OCP:CHANnel:TRIGger

COMMAND DETAILS 27 *ESE Status Command Description The Standard Event Status Enable command determines which events in the Standard Event Status

Pagina 92 - :OCP:STATus?

PEL-2000 Programming Manual 28 Return Parameter <NR1> Bit(s) Set <NR1> Bit(s) Set 4 QYE 32 CME 8 DDE 64 ~ 16 EXE 128 ~ Example *ESR

Pagina 93 - :OCP:RUN

COMMAND DETAILS 29 Query Example *OPC? 1 All pending operations are completed. *RCL Status Command Description The Recall Instrument State comm

Pagina 94 - Program Subsystem

TABLE OF CONTENTS 3 Table of Contents INTERFACE OVERVIEW ... 5 Rear Panel Overview ...

Pagina 95 - :PROGram:MEMory

PEL-2000 Programming Manual 30 *SAV All Channels Description Saves the data memory into the specified save slot. Syntax *SAV <NR1> Parame

Pagina 96 - (BIT WEIGHT)

COMMAND DETAILS 31 Example *SRE? 48 Returns settings of the Service Request Enable Register. Here ESB and MAV are returned. *STB? Status Command

Pagina 97 - :PROGram:SEQuence:MODE

PEL-2000 Programming Manual 32 Abort Subsystem :ABORt ... 32 :ABORt Al

Pagina 98 - :PROGram:ACTive

COMMAND DETAILS 33 Channel Subsystem :CHANnel[:LOAD] ... 33 :CHANnel:ACTive ...

Pagina 99 - :PROGram:ONTime

PEL-2000 Programming Manual 34 :CHANnel:ACTive Channel Specific Command Description This command is for compatibility with other instruments on

Pagina 100 - :PROGram:OFFTime

COMMAND DETAILS 35 :CHANnel:SYNCon:ALL All Channels Description Turns independent mode on or off for all the channels. Syntax :CHANnel:SYNCon:ALL

Pagina 101 - :PROGram:PFTime

PEL-2000 Programming Manual 36 MAX Last channel MIN First channel Example :CHAN:DISP 1 Sets to the active channel on the display to ch1. Query S

Pagina 102 - :PROGram:CHAin:STARt

COMMAND DETAILS 37 :MEMo? Channel Specific Query Description Creates or returns the “memo” that is displayed in the “System Information” screen i

Pagina 103 - Resistance Subsystem

PEL-2000 Programming Manual 38 CONFIGURE Subsystem :CONFigure:VOLTage:ON ... 38 :CONFigure:VOLTage:RANG

Pagina 104

COMMAND DETAILS 39 Example :CONF:VOLT:ON 30MV Set Von to 30 millivolts. Query Syntax :CONFigure:VOLTage:ON? Return Parameter <NR2> I unit =

Pagina 105 - :RESistance:STATic:RECall

PEL-2000 Programming Manual Power Subsystem ... 135 SYSTEM Subsystem ... 141 Memory Su

Pagina 106

PEL-2000 Programming Manual 40 :CONFigure:VOLTage:LATch Channel Specific Command Description Turn Von Latch on or off for the specific channel.

Pagina 107

COMMAND DETAILS 41 Query Example :CONF:AUTO:LOAD? 1 Auto load is On. :CONFigure:AUTO:MODE All channels Description Configures the Auto Load mode

Pagina 108

PEL-2000 Programming Manual 42 Return Parameter <NR1> SOUND Status 0 Off 1 On Query Example :CONF:SOUND? 0 Sound is off for the specific c

Pagina 109

COMMAND DETAILS 43 Parameter OLD/0 Old UPDATED/1 Updated Example :CONF:LOAD OLD Configuration type set as OLD. Query Syntax :CONFigure:LOAD? Retu

Pagina 110

PEL-2000 Programming Manual 44 :CONFigure:PROTection:CURRent:LEVel Channel Specific Command Description Sets the current protection level for the

Pagina 111

COMMAND DETAILS 45 :CONFigure:PROTection:VOLTage:STATe Channel Specific Command Description Sets the voltage protection for the current/specific

Pagina 112

PEL-2000 Programming Manual 46 MIN Sets to the minimum level MAX Sets the voltage limit to the maximum level Example :CONF:PROT:VOLT:LEV MAX Sets

Pagina 113

COMMAND DETAILS 47 :CONFigure:PROTection:POWer:LEVel Channel Specific Command Description Sets the power protection level for the current/specifi

Pagina 114

PEL-2000 Programming Manual 48 :CONFigure:PROTection:UVP:LEVel Channel Specific Command Description Sets the under voltage protection level for t

Pagina 115 - RUN Subsystem

COMMAND DETAILS 49 Example :CONF:RESP 0 Response set to normal. Query Syntax : CONFigure:RESPonse? Return Parameter <NR1> Response 0 Norma

Pagina 116 - SHOW Subsystem

INTERFACE OVERVIEW 5 INTERFACE OVERVIEW This manual describes how to use the PEL-2000’s remote command functionality and lists the command details

Pagina 117

PEL-2000 Programming Manual 50 Query Example : CONF:GRO:UNIT? 2 2 units are set for the parallel mode. :CONFigure:GROup:MODE Channel Specific Comm

Pagina 118 - SPECIFICATION Subsystem

COMMAND DETAILS 51 Utility Subsystem :UTILity:AUTO:LOAD ... 51 :UTILity:AUTO:MODE ...

Pagina 119

PEL-2000 Programming Manual 52 :UTILity:AUTO:MODE System Command Description Sets the mainframe auto mode as load or program. Upon startup the ma

Pagina 120

COMMAND DETAILS 53 1 On Query Example :UTIL:SOUN? 0 Sound is currently set to off. :UTILity:REMote System Command Description Turns the remote c

Pagina 121 - :SPECification:TEST:

PEL-2000 Programming Manual 54 :UTILity:TIME System Command Description Sets the date and time on the mainframe. Syntax :UTILity:TIME [aard] Par

Pagina 122 - STATUS Subsystem

COMMAND DETAILS 55 0 Old 1 Updated Query Example :UTIL:LOAD? 1 The knob style is set to Updated. :UTILity:IDENtify System Command Description F

Pagina 123 - :STATus:CHANnel:EVENt?

PEL-2000 Programming Manual 56 Current Subsystem :CURRent:STATic:RECall ... 56 :CURRent:STATic:L1/L2

Pagina 124 - PTRansition

COMMAND DETAILS 57 Query Example :CURR:STAT:REC? 0 A Value is active. :CURRent:STATic:L1/L2 Channel Specific Command Description Sets the A/B Val

Pagina 125 - :STATus:CSUMmary:ENABle

PEL-2000 Programming Manual 58 Query Example :CURR:STAT:L2? 2 Returns the current setting (2 A) for B Value. :CURRent:STATic:RISE/FALL Channel Sp

Pagina 126 - :STATus:CSUMmary:EVENt?

COMMAND DETAILS 59 :CURRent:STATic:LOW:AVALue/BVALue Channel Specific Command Description Sets the low range A/B Value for constant current stati

Pagina 127 - :STATus:QUEStionable:ENABle

PEL-2000 Programming Manual 6 Configuring the USB Interface USB connection PC side connector Type A, host PEL-2000 side connector Type B, device

Pagina 128

PEL-2000 Programming Manual 60 :CURRent:STATic:LOW:RISE/FALL Channel Specific Command Description Sets the low range rising/falling slew rates.

Pagina 129 - :STATus:PREset

COMMAND DETAILS 61 AVALue 10 Sets A Value to 10 Amps. (high range only) BVALue 20 Sets B Value to 20 Amps. (high Range only) AVALue MIN Sets A

Pagina 130 - Voltage Subsystem

PEL-2000 Programming Manual 62 RISE/FALL MIN Sets to the slowest rising/falling slew rate. RISE/FALL MAX Sets to the fastest rising/falling slew

Pagina 131 - :VOLTage:AVALue/BVALue

COMMAND DETAILS 63 L1/L2 MAX Sets L1 or L2 to the maximum Level for the specific channel. Example :CURR:DYN:L1 10 In CC dynamic mode, Set L1 (lev

Pagina 132 - :VOLTage:CURRent

PEL-2000 Programming Manual 64 Return Parameter <NR2> [MAX|MIN] Slew rate 1 Unit=1 amp/uS Returns the slew rate. MAX/MIN Returns the allow

Pagina 133 - :VOLTage:MODE

COMMAND DETAILS 65 Query Example :CURR:DYN:LOW:T1? 2.5 :CURR:DYN:LOW:T1? MIN 0.000025 Returns the T1 time of 2.5 seconds. Returns the minimum T1 t

Pagina 134

PEL-2000 Programming Manual 66 MAX/MIN Returns the allowable maximum and minimum. Query Example :CURR:DYN:LOW:L2? 2.0400 Returns current for the

Pagina 135 - Power Subsystem

COMMAND DETAILS 67 :CURRent:DYNamic:LOW:T1/T2 Channel Specific Command Description Sets the low range timers T1 or T2 for CC dynamic mode for the

Pagina 136 - :POWer:CURRent

PEL-2000 Programming Manual 68 Syntax :CURRent:DYNamic:HIGH:L1/L2 <NRf+>[A] Parameter NRf+[A] L1 10 Sets L1 to 10 Amps. (High range only)

Pagina 137 - :POWer:LOW:AVALue/BVALue

COMMAND DETAILS 69 RISE/FALL 1 Sets the rising/falling slew rate to 1A/uS RISE/FALL MIN Sets to the slowest rising/falling slew rate. RISE/FALL

Pagina 138 - :POWer:LOW:CURRent

INTERFACE OVERVIEW 7 7. On the PC, activate a terminal application such as MTTTY (Multi-Threaded TTY). To check the COM port No., see the Device

Pagina 139 - :POWer:HIGH:AVALue/BVALue

PEL-2000 Programming Manual 70 Example :CURR:DYNA:HIGH:T1 10S Sets the high range T1 time to 10 seconds for the specific channel. Query Syntax :CU

Pagina 140 - :POWer:HIGH:CURRent

COMMAND DETAILS 71 FETCH Subsystem :FETCh:VOLTage? ... 71 :FETCh:CURRent? ...

Pagina 141 - SYSTEM Subsystem

PEL-2000 Programming Manual 72 :FETCh:POWer? Channel Specific Status Command Description This query returns the real-time power of the load modul

Pagina 142 - :SYSTem:SETup

COMMAND DETAILS 73 Parameter <aard> CH1,CH2,CH3,CH4,CH5,CH6,CH7,CH8 Returns all the voltage values from all the channels, 1-8(PEL-2004)/1-

Pagina 143 - Memory Subsystem

PEL-2000 Programming Manual 74 Parameter <aard> CH1,CH2,CH3,CH4,CH5,CH6,CH7,CH8 Returns all the power values from all the channels, 1-8(PE

Pagina 144 - :MEMory:RECall:PREset

COMMAND DETAILS 75 LOAD Subsystem :LOAD[:STATe] ... 75 :LOAD:SHORt[:STATe] ...

Pagina 145 - :MEMory:RECall:SETup

PEL-2000 Programming Manual 76 Parameter ON/1 Shorting is On OFF/0 Shorting is Off Example :LOAD:SHOR ON Short circuits the load module channel.

Pagina 146 - :MEMory:FILE:PROGram

COMMAND DETAILS 77 Description Returns the protection levels for electronic load Query Syntax :LOAD:PROTection? Query Example :LOAD:PROT? Clears t

Pagina 147 - :MEMory:FILE:SEQuence

PEL-2000 Programming Manual 78 Query Example :LOAD:TIME? 5.1 Returns the load on time as 5.1 seconds. :LOAD:DELay Channel Specific Command Desc

Pagina 148

COMMAND DETAILS 79 Query Example :LOAD:TYPE? 0 The normal load type is active

Pagina 149 - SEQuence Subsystem

PEL-2000 Programming Manual 8 RS-232C Interface Configuration RS-232C configuration Connector DB-9, Male Baud rate 2400, 4800, 9600, 19200, 38400

Pagina 150 - :SEQuence:POINt:RESistance

PEL-2000 Programming Manual 80 Measure Subsystem :MEASure:VOLTage? ... 80 :MEASure:CURRent? ..

Pagina 151 - :SEQuence:POINt:RISE/FALL

COMMAND DETAILS 81 :MEASure:POWer? Channel Specific Command Description This query returns the measured power of the specific channel. Query Sy

Pagina 152

PEL-2000 Programming Manual 82 :MEASure:SCAN Channel Specific Command Description This command allows the mainframe to scan all the load modules.

Pagina 153 - :SEQuence:END:LOAD

COMMAND DETAILS 83 Query Example :MEAS:ALLV? 2.5000, 3.0000, 0.0000, 0.0000, 0.0000, 0.0000, 5.500, 0.0000 Channel 1 and 2 have voltages of 2.5 an

Pagina 154 - :SEQuence:VOLTage:RANGe

PEL-2000 Programming Manual 84 Query Example :MEAS:ALLP? 0.0000, 0.0000, 0.0000 ,0.0000, 0.0000, 0.0000, 1.5000, 3.2000 Channels 1 to 6 have no po

Pagina 155 - :SEQuence:CHANnel:TIME

COMMAND DETAILS 85 MODE Subsystem :MODE ... 85 :MODE Channel Specific

Pagina 156 - :SEQuence:TRIGger:IN

PEL-2000 Programming Manual 86 CCDL CC dynamic mode, low range CCDH CC dynamic mode, high range CRL CR static mode, low range CRH CR static mo

Pagina 157 - :SEQuence:TRIGger:IN:CHANnel

COMMAND DETAILS 87 OCP Test Automation Commands :OCP:EDIT:CHANnel? ... 87 :OCP:CHANnel:RANGe ...

Pagina 158

PEL-2000 Programming Manual 88 :OCP:CHANnel:RANGe Channel Specific Command Description Sets or queries the channel range. High(CC Mode High) or L

Pagina 159 - GLOBal Subsystem

COMMAND DETAILS 89 :OCP:CHANnel:END Channel Specific Command Description Sets the ending current value for the test. The value must be higher tha

Pagina 160

INTERFACE OVERVIEW 9 5. Edit the Baud rate, Stop bit and parity. Baud rate 2400, 4800, 9600, 19200, 38400 Stop Bit Range 1,2 Parity Range

Pagina 161 - Command Error Codes

PEL-2000 Programming Manual 90 :OCP:CHANnel:LAST Channel Specific Command Description Queries or sets the current value for after the DUT OCP pr

Pagina 162 - STATUS REGISTERS

COMMAND DETAILS 91 :OCP:CHANnel:DELay Channel Specific Command Description Queries or sets the test delay time for the OCP Test Automation functi

Pagina 163 - STATUS REGISTERS

PEL-2000 Programming Manual 92 :OCP:CHANnel:ACTive Channel Specific Command Description Queries or sets which bit(s) are the active channel for t

Pagina 164 - Channel Status

COMMAND DETAILS 93 1 OCP test active Query Example :OCP:STAT? 0 The test has ended. :OCP:SAVE Channel Spe

Pagina 165 - Channel Summary

PEL-2000 Programming Manual 94 Program Subsystem :PROGram:FILE ... 94 :PROGram:SEQuenc

Pagina 166 - Questionable Status

COMMAND DETAILS 95 :PROGram:SEQuence Program Number Specific Description Sets the Sequence number for the current program number. Syntax :PROGra

Pagina 167 - Data inData out

PEL-2000 Programming Manual 96 Query Example :PROG:MEM? 1 The memory number for the current program/sequence is 001. :PROGram:SEQuence:SHORt:CHAN

Pagina 168 - Standard Event Status

COMMAND DETAILS 97 :PROGram:SEQuence:SHORt:TIME Program Number Specific Description Sets the short time (seconds) for the current program sequenc

Pagina 169 - Status Byte Register

PEL-2000 Programming Manual 98 SKIP Skip mode: current program sequence is skipped. Example :PROG:SEQ:MODE: AUTO The current program sequence is

Pagina 170 - Service Request Register

COMMAND DETAILS 99 Query Example :PROG:ACT? 12 Channels 3 and 4 are active. :PROGram:CHAin Program Number Specific Description Chains the curre