COMMAND DETAILS
131
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:VOLT:L1?
5
:VOLT:L1? MAX
81.6000
A Value is set to 5 volts.
Returns the maximum
settable voltage.
:VOLTage:RECall
Channel Specific
Command
Description
Sets or queries whether A Value or B Value is the
currently active value in CV mode.
Syntax
:VOLTage:RECall {A|0|B|1}
Parameter
A/0
A
B/1
B
Example
:VOLT:REC 1
Makes B Value the active
value.
Query Syntax
:VOLTage:RECall?
Return Parameter
<NR1>
Value
0
A
1
B
:VOLTage:AVALue/BVALue
Channel Specific
Command
Description
Sets the voltage of A Value or B Value in CV mode.
Syntax
:VOLTage:AVALue/BVALue <NRf+>[V]
Parameter
<NRf+>[V]
Voltage
10
10 volts
10V
10 volts
MIN
Sets the voltage to the
minimum value for the
channel