COMMAND DETAILS
29
Query Example
*OPC?
1
All pending operations
are completed.
*RCL
Status Command
Description
The Recall Instrument State command restores the
instrument settings from a previously saved
memory setting.
Syntax
*RCL <NRf>
Parameter
<NRf>
Recall Memory Setting
1~120
1~120
Example
*RCL 1
Recalls Setting memory 1
*RDT?
System Command
Description
Returns the load module type in each channel in
order from 1~8. If no frame is present a 0 is
returned.
Query Syntax
*RDT?
Return Parameter
<aard>
Occupied Channel
2020L
PEL-2020 left channel
0
Empty channel
Query Example
*RDT?
0,0,2020L,2020R,0,0,0,0
Channels 1-2 and 5-8 are
empty; 3-4 is occupied
by the PEL-2020 load
module.
*RST
Status Command
Description
Resets the mainframe by forcing the ABORt, *CLS,
and LOAD:PROT:CLE command.
Syntax
*RST
Example
*RST
Vedere la pagina 28